Tip 129: ADVANTEC® System: Ultra MAXVAC®
The ADVANTEC® Alcon LEGACY®
system permits the surgeon to set a maximum vacuum level of "greater
than 500 mm Hg" ("500+" on the LEGACY screen). With this
vacuum setting, the pump will continue to run regardless of the vacuum
level achieved.
The use of Ultra MAXVAC® will permit nuclei of grade I
or II density to be aspirated without the use of any ultrasonic energy.
There are 2 strict requirements, however.
- The "greater than 500 mm Hg" setting should only be used
with the 1.1 FLARED tip and a relatively tight incision (I prefer the
angulated Mackool* 1.1 ABS® FLARED tip and a 2.75 mm incision).
- The infusion bottle height (drip chamber to eye level) should be at
least 110 cm.
|

|